Human Myostatin,MSTN ELISA Kit
-
中文名称:
人肌抑素(MSTN)酶联免疫试剂盒
-
货号:
CSB-E11300h
-
规格:
96T/48T
-
价格:
¥3600/¥2500
-
其他:
产品详情
-
产品描述:
This Human MSTN ELISA Kit was designed for the quantitative measurement of Human MSTN protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 0.625 ng/mL-20 ng/mL and the sensitivity is 0.312 ng/mL.
-
别名:
Cmpt ELISA Kit; GDF-8 ELISA Kit; GDF8_HUMAN ELISA Kit; Growth/differentiation factor 8 ELISA Kit; MSLHP ELISA Kit; MSTN ELISA Kit; MSTN myostatin ELISA Kit; Myostatin ELISA Kit; Myostatin Propeptide ELISA Kit
-
缩写:
-
Uniprot No.:
-
种属:
Homo sapiens (Human)
-
样本类型:
serum, plasma, tissue homogenates
-
检测范围:
0.625 ng/mL-20 ng/mL
-
灵敏度:
0.312 ng/mL
-
反应时间:
1-5h
-
样本体积:
50-100ul
-
检测波长:
450 nm
-
研究领域:
Signal Transduction
-
测定原理:
quantitative
-
测定方法:
Sandwich
-
精密度:
Intra-assay Precision (Precision within an assay): CV%<15% |
|
|
|
Three samples of known concentration were tested twenty times on one plate to assess. |
|
Inter-assay Precision (Precision between assays): CV%<15% |
|
|
|
Three samples of known concentration were tested in twenty assays to assess. |
|
|
|
|
|
|
|
|
|
-
线性度:
To assess the linearity of the assay, samples were spiked with high concentrations of human MSTN in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
|
Sample |
Serum(n=4) |
|
1:1 |
Average % |
95 |
|
Range % |
92-99 |
|
1:2 |
Average % |
94 |
|
Range % |
91-98 |
|
1:4 |
Average % |
101 |
|
Range % |
97-105 |
|
1:8 |
Average % |
86 |
|
Range % |
82-90 |
|
-
回收率:
The recovery of human MSTN spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
Sample Type |
Average % Recovery |
Range |
|
Serum (n=5) |
93 |
87-97 |
|
EDTA plasma (n=4) |
86 |
82-91 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-
标准曲线:
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
ng/ml |
OD1 |
OD2 |
Average |
Corrected |
|
20 |
1.851 |
1.992 |
1.922 |
1.878 |
|
10 |
1.058 |
1.136 |
1.097 |
1.053 |
|
4.375 |
0.502 |
0.524 |
0.513 |
0.469 |
|
1.875 |
0.268 |
0.255 |
0.262 |
0.218 |
|
0.625 |
0.116 |
0.119 |
0.118 |
0.074 |
|
0 |
0.043 |
0.045 |
0.044 |
|
|
|
-
本试剂盒所含材料:
- A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-human MSTN antibody.
- Six vials standard (5 ml/bottle) ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
- One vial HRP-conjugated MSTN antibody (6 ml/bottle) ---Bind to the MSTN in the samples or standards and react with the substrate to make the solution chromogenic.
- Substrate A (1 x 7 ml)
- Substrate B (1 x 7 ml)
- One vial Stop Solution (7ml/bottle) ---Stop the color reaction. The solution color immediately turns from blue to yellow.
- Four Adhesive Strips (For 96 wells) ---Cover the microplate when incubation.
- An instruction manual
显示更多
收起更多
-
本试剂盒不含材料:
- A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 600 nm or 630 nm.
- An incubator can provide stable incubation conditions up to 37°C±5°C.
- Centrifuge
- Vortex
- Squirt bottle, manifold dispenser, or automated microplate washer
- Absorbent paper for blotting the microtiter plate
- 50-300ul multi-channel micropipette
- Pipette tips
- Single-channel micropipette with different ranges
- 100ml and 500ml graduated cylinders
- Deionized or distilled water
- Timer
- Test tubes for dilution
显示更多
收起更多
-
数据处理:
-
货期:
3-5 working days
引用文献
-
Novel Approach for Glycemic Management Incorporating Vibration Stimulation of Skeletal Muscle in Obesity
M Kim,International Journal of Environmental Research and Public Health,2023
-
Effects of functional training with blood occlusion on the irisin, follistatin, and myostatin myokines in elderly men
F Pazokian,European Review of Aging and Physical Activity,2022
-
Strategic Ingestion of High-Protein Dairy Milk during a Resistance Training Program Increases Lean Mass, Strength, and Power in Trained Young Males
Maryam Pourabbas,Nutrients,2021
-
Comparison of whole egg v. egg white ingestion during 12 weeks of resistance training on skeletal muscle regulatory markers in resistance-trained men
R Bagheri,Br J Nutr,2020
-
Urinary Levels of titin-n fragment, a Skeletal Muscle Damage Marker, are increased in Subjects with nonalcoholic fatty Liver Disease
Oshida N,Scientific Reports,2019
-
Blood Flow Restriction During Futsal Training Increases Muscle Activation and Strength
Sadegh Amani-Shalamzari,Frontiers in Physiology,2019
-
High-Intensity Aerobic Exercise Improves Both Hepatic Fat Content and Stiffness in Sedentary Obese Men with Nonalcoholic Fatty Liver Disease
Sechang Oh,Sci Rep. ,2017
-
Serum concentration of myostatin, insulin and glucose response to continuous and discontinuous exercises in overweight menopausal and menstrual women
Masomeh Hannani.et al,Hormozgan Medical Journal,2016
-
Myostatin and carbohydrate disturbances
Assyov YS.et al,Endocr Res.,2016
-
The correlation of resistance exercise-induced myostatin with insulin resistance and plasma cytokines in healthy young men
Kazemi F,J Endocrinol Invest,2015
Most popular with customers